Click to enlargeRacal 1260-35 VXI High-density Scanner/Multiplexer w/ option 001 Smart Card

DC PERFORMANCE
Path Resistance <1.0W (1x96 configuration) <0.5W (1x6 configuration)
Isolation >7.5 x 108W
AC PERFORMANCE (into 50W)
Capacitance Open Channel: <600 pF (1x96 configuration) Channel-Chassis: <200 pF (1x96 configuration) High-Low: <600 pF (1x96 configuration)
Bandwidth(-3dB) >15 MHz (1x48 configuration) >50 MHz (1x6 configuration)
Insertion Loss 100kHz: < 0.1 dB (1x6 configuration) 1MHz: < 0.5 dB (1x6 configuration) 10MHz: < 1.0 dB (1x6 configuration)
Crosstalk 100kHz: <-90 dB 1MHz: <-70 dB
Cooling Requirements Airflow: 4.0 liters/Sec
Back Pressure: 0.5 mm H20
Power Requirements (Ipm) +5V: 0.4A (2.8A with Option 01 installed) +24V: 10 mA per relay (energized)
Programming Syntax is in the form: “.
Example: CLOSE 3.02
This CLOSE statement will close channel number 2 on the 1260-35 module at card address 3.

Manual in PDF format.

CURRENTLY SOLD OUT


Rac-1260-35-CAP-GR2

Unless specifically stated, manuals and/or accessories are not included